Gas Blending

Precision Gas Blending & Integrated Monitoring:

The ultra-compact Quantum Blending System employs the very latest gas blending technologies to ensure the precision delivery of FiO2 (21 to 100%) and sweep flow rates. In its standard configuration the Quantum Blending System provides a sweep flow range from 0.025 to 10 L/m when flowing a single gas delivery and up to a maximum flow rate of  20 L/m when modulating the delivery of at least two gases.

The Blending System is further enhanced with an additional Precision Low Flow control module specifically designed to support the needs of the pediatric patient population.

Included as standard is an independent monitoring system that continuously monitors measured values with the expected outputs from the gas blending system. The fault detection strategy is based on the detection of a discrepancy between the monitoring and blending system, a performance alarm is triggered when the measured discrepancy exceeds 10%.